 ===== CytoFLEX LX - Flow Cytometer ===== ===== CytoFLEX LX - Flow Cytometer =====
  
-{{:​aree:​citometria:​strumentazione:​cytoflex_lx.jpeg?​300 }} The CytoFLEX LX Flow Cytometer ​is an expansion of the CytoFLEX ​Platform. This model expands research possibilities ​with up to six lasers ​and 21 color parameters. Six spatially separated lasers allows panels to be spread across the spectrum reducing cross talk and spectral overlap.+{{:​aree:​citometria:​strumentazione:​cytoflex_lx.jpeg?​300 }} The CytoFLEX LX flow cytometer ​is an expansion of the CytoFLEX ​platform: a unique instrument based on photodiode technology, equipped ​with 5 lasers, capable of detecting ​up to 19 fluorescences ​and 2 scattering ​parameters ​simultaneously.
  
-For higher throughput applications an optional plate loader module is available.+Digital signal processing on 7 decades of dynamic range. 
 + 
 +Excellent resolution of nanoparticles thanks to the new VSSC parameter. 
 + 
 +\\ \\ 
 + 
 +==== Technical specifications ==== 
 + 
 +  * Resolution:​ 
 +    * < 300 nm particles (488 nm side scatter) 
 +    * < 80 nm polystyrene particles (405 nm side scatter) 
 +    * rCV 3% 
 +  * collections vessels: 
 +    * 5 ml, 2 ml, 1.5 ml; 
 +    * 96w plates; 
 +  * absolute counting.
  
 \\ \\
  
-  * {{ :aree:​citometria:​strumentazione:​cytoflex_lx-spec-ita-rev2.pdf |Technical specifications}} +==== Optical configuration ==== 
-  * {{ :​en:​aree:​citometria:​strumentazione:​cytoflex_lx-opt_conf.pdf ​|Optical configuration}}+ 
 +5 activated lasers / 19 color parameters
 + 
 + **UV Laser** \\ <wrap lo>355 nm</​wrap> ​   ^  **Violet Laser** \\ <wrap lo>405 nm</​wrap> ​      ​^ ​ **Blue Laser** \\ <wrap lo>488 nm</​wrap> ​ ^  **Yellow-Green Laser** \\ <wrap lo>561 nm</​wrap> ​ ^  **Red Laser** \\ <wrap lo>640 nm</​wrap> ​ ^ 
 + ​405/​30 \\ 525/40 \\ 675/​30 ​               |  450/45 \\ 525/40 \\ 610/20 \\ 660/10 \\ 763/​43 ​  ​| ​ 488/8 SSC \\ 525/40 \\ 610/20 \\ 690/​50 ​  ​| ​ 585/42 \\ 610/20 \\ 675/30 \\ 710/50 \\ 763/​43 ​   |  660/10 \\ 712/25 \\ 763/​43 ​              | 
  
 \\ \\ \\ \\ \\ \\
Line 33: Line 53:
 ===== FACSCanto - BD Biosciences ===== ===== FACSCanto - BD Biosciences =====
  
-{{:​aree:​citometria:​strumentazione:​facscanto.png?​300 |}}  \\ \\ **Technical features (FIXME brochure):​**+{{:​aree:​citometria:​strumentazione:​facscanto.png?​300 |}}  \\ \\ **Technical features ({{:​aree:​citometria:​strumentazione:​facscanto-brochure.pdf|brochure}}):**
   * 2 laser (6 fluorescence)   * 2 laser (6 fluorescence)